"Notes:\n"
"\n"
"(1) If the primary measurement function is diode or continuity, this "
"attribute returns the maximum forward voltage of a junction at "
"approximately 0.7 mA.\n"
2 all "Secondary Auto Range Value" FL45_ATTR_SECONDARY_AUTO_RANGE_VALUE ViReal64 g
"Always returns the actual range the DMM is currently using for the "
"secondary measurement. The secondary measurement function is always "
"auto-ranging. The value is in units appropriate for the "
"FL45_ATTR_SECONDARY_FUNCTION attribute as shown below.\n"
" \n"
" DC Volts - volts\n"
" AC Volts - volts\n"
" DC Current - amperes\n"
" AC Current - amperes\n"
" 2-Wire Resistance - ohms\n"
" Frequency - hertz\n"
" Diode - volts\n"
"\n"
"(1) If the secondary measurement function is diode, this attribute returns "
"the maximum forward voltage of a junction at approximately 0.7 mA\n"
2 all "Resolution" FL45_ATTR_RESOLUTION ViReal64 gs attrResolutionRangeTable
"Specifies the measurement resolution in digits of precision.\n"
" Setting this attribute to higher values increases the measurement "
"accuracy. Setting this attribute to lower values increases the "
"measurement speed. I\n"
"\n"
"Notes:\n"
"\n"
"(1) If the secondary measurement function is enabled, then this attribute "
"also specifies the resolution of the secondary measurement.\n"
2 all "Trigger Source" FL45_ATTR_TRIGGER_SOURCE ViInt32 gs attrTriggerSourceRangeTable
"Specifies the trigger source. After you call one of the FL45_Measure, "
"FL45_Read, or FL45_Initiate functions, the DMM waits for the trigger you "
"specify with this attribute. After it receives the trigger, the DMM waits "
"the length of time you specify with the FL45_ATTR_TRIGGER_DELAY attribute. "
" The DMM then takes a measurement.\n"
2 all "Trigger Delay" FL45_ATTR_TRIGGER_DELAY ViReal64 gs
"This attribute specifies the value you want to use for the trigger delay. "
"Express this value in seconds.\n"
"\n"
"The trigger delay specifies the length of time the DMM waits after it "
"receives the trigger and before it takes a measurement.\n"
"\n"
"The FL45 allows only auto-delay enabled or no delay at all. You cannot "
"set the FL45 to a manual delay value other than 0.0. The driver reserves "
"negative values for configuring the DMM to calculate the trigger delay "
"automatically.\n"
"\n"
"Defined Constant Values:\n"
"FL45_VAL_AUTO_DELAY_ON - Auto-delay on\n"
"\n"
"Valid Manual Range: 0.0 (seconds)\n"
"\n"
"Default Value: FL45_VAL_AUTO_DELAY_ON\n"
"\n"
"Notes:\n"
"\n"
"(1) Setting this parameter to FL45_VAL_AUTO_DELAY_ON configures the DMM to "
"calculate the best trigger delay before each measurement.\n"
2 all "AC Minimum Frequency" FL45_ATTR_AC_MIN_FREQ ViReal64 gs
"Specifies the minimum frequency component of the input signal for AC "
"measurements. The value of this attribute affects instrument behavior only "
"when you set the FL45_ATTR_FUNCTION attribute to an AC measurement.\n"
"\n"
"Valid Range: The valid range depends on AC measurement as shown below.\n"
"\n"
"AC Volts: 20.0 to 100.0E3 Hz\n"
"AC Current: 20.0 to 20.0E3 Hz\n"
"AC+DC Volts: 20.0 to 100.0E3 Hz\n"
"AC+DC Current: 20.0 to 20.0E3 Hz\n"
"\n"
"When you set the Measurement Function to AC Current or to AC+DC Current "
"and set the Range to a value greater than 0.1 A, then the maximum value "
"the DMM allows for this parameter decreases to 2.0 kHz.\n"
"\n"
"Notes:\n"
"\n"
"(1) The driver coerces all valid values you pass for this parameter to "
"20.0 Hz.\n"
2 all "AC Maximum Frequency" FL45_ATTR_AC_MAX_FREQ ViReal64 gs
"Specifies the maximum frequency component of the input signal for AC "
"measurements. The value of this attribute affects instrument behavior only "
"when you set the FL45_ATTR_FUNCTION attribute to an AC measurement.\n"
"\n"
"Valid Range: The valid range depends on AC measurement function as shown "
"below.\n"
"\n"
"AC Volts: 20.0 to 100.0E3 Hz\n"
"AC Current: 20.0 to 20.0E3 Hz\n"
"AC+DC Volts: 20.0 to 100.0E3 Hz\n"
"AC+DC Current: 20.0 to 20.0E Hz\n"
"\n"
"When you set the Measurement Function to AC Current or to AC+DC Current "
"and set the Range to a value greater than 0.1 amperes, then the maximum "
"value the DMM allows for this parameter decreases to 2.0 kHz.\n"
"\n"
"Notes:\n"
"\n"
"(1) The driver coerces all valid values you pass for this parameter to the "
"maximum frequency.\n"
2 all "Aperture Time" FL45_ATTR_APERTURE_TIME ViReal64 g
"Returns the measurement aperture time for the current configuration. The "
"units for this attribute are seconds.\n"
2 all "Aperture Time Units" FL45_ATTR_APERTURE_TIME_UNITS ViInt32 g
"Returns the units for the attribute FL45_ATTR_APERTURE_TIME. Aperture "
"time is always given in units of seconds.\n"
"\n"
"Possible values are the following:\n"
"FL45_VAL_SECONDS\n"
" "
1 all "Advanced Triggering"
"Attributes for acquiring data on multiple triggers and multiple "
"measurements per trigger.\n"
" "
1 all "Miscellaneous"
"Miscellaneous attributes.\n"
" "
1 all "Function Modifiers"
"Attributes that control the FLuke 45 function modifiers. A function "
"modifier causes the meter to modify the normal operation of a measurement "
"function (or perform an action on a measurement) before displaying a "
"reading. For example, the relative modifier causes the meter to display "
"the difference between a measured value and the relative base."
2 all "Decibel Modifier"
"This group specifies attributes for turning on/off and configuring a "
"decibel modifier. A modifier causes the DMM to modify the normal "
"operation of a measurement function (or performs an action on a "
"measurement) before displaying a reading.\n"
3 all "dB Mode" FL45_ATTR_DB_MODE ViInt32 gs attrDbModeRangeTable
"Specifies the decibel modifier to use. The DMM must be in a voltage "
"measurement function before this attribute can be set to a decibel "
"modifier.\n"
"\n"
"When you enable a decibel modifier, the readings you obtain from the "
"FL45_Measure, FL45_Read, and FL45_Fetch functions are in the decibel "
"units you configure with this function.\n"
"\n"
"Notes:\n"
"\n"
"(1) You can use the decibel modifiers for only voltage measurement "
"functions.\n"
3 all "dB Reference Impedance" FL45_ATTR_DB_REF ViInt32 gs attrDbRefRangeTable
"Specifies the reference impedance used for the decibel modifier.\n"
"\n"
"If you set the Decibel Mode to dB Power, the valid impedance values are 2, "
"4, 8, or 16 ohms.\n"
"\n"
"Note:\n"
"\n"
"(1) This parameter affects instrument behavior only when you enable one of "
"the decibel modifier's calculation modes.\n"
2 all "Hold Modifier"
"This group contains attributes that control the DMM's hold modifier. The "
"hold modifier configures the DMM to take a new measurement only when the "
"input signal is stable and 'hold' that measurement on the display. This "
"feature can be particularly advantageous in difficult or hazardous "
"circumstances when you might want to keep your eyes fixed on the probes, "
"and then read the display when it is safe or convenient to do so."
3 all "Hold Enable" FL45_ATTR_HOLD_ENABLE ViBoolean gs Boolean_values
"Specifies whether to enable the hold modifier. VI_TRUE enables the "
"modifier, and VI_FALSE disables the modifier. The hold modifier "
"configures the DMM to take a new measurement only when the input signal is "
"stable and 'hold' the measurement on the display. This feature can be "
"particularly advantageous in difficult or hazardous circumstances when you "
"might want to keep your eyes fixed on the probes, and then read the "
"display when it is safe or convenient to do so.\n"
"\n"
"Notes:\n"
"\n"
"(1) Use the FL45_Hold() function to force a new reading to be displayed. "
"When a new, stable reading is detected, a beep is emitted, and the display "
"is automatically updated.\n"
"\n"
"(2) The FL45_Hold() function should be called before the trigger occurs. "
"For example if you are using the software triggering mechanism. Then the "
"order of function calls should be:\n"
"\n"
" FL45_Initiate ();\n"
" FL45_Hold ();\n"
" FL45_SendSWTrigger ();\n"
3 all "Hold Threshold" FL45_ATTR_HOLD_THRESHOLD ViInt32 gs attrHoldThresholdRangeTable
"Specifies the hold measurement threshold. The DMM takes a new measurement "
"only when the input signal is as stable as you specify with this "
"attribute.\n"
"\n"
"Notes:\n"
"\n"
"(1) This attribute affects instrument behavior only when you set the "
"FL45_ATTR_HOLD_ENABLE attribute to VI_TRUE.\n"
" "
2 all "Compare Modifier"
"This group specifies attributes for turning on/off and configuring the "
"compare modifier. The compare provides an easy way to determine if a "
"reading falls within a designated range of values.\n"
3 all "Compare Enable" FL45_ATTR_COMPARE_ENABLE ViBoolean gs Boolean_values
"Specifies whether to enable the compare modifier. VI_TRUE enables the "
"compare modifier, and VI_FALSE disables the compare modifier.\n"
"\n"
"The compare modifier provides an easy way to determine if a reading falls "
"within a range you specify. You specify the range with the High Value and "
"Low Value parameters. Enabling the compare modifier automatically enables "
"the hold modifier. Disabling the hold does not affect the compare "
"modifier.\n"
"\n"
"Use the FL45_FetchCompare function to obtain the comparison result.\n"
3 all "Compare Low" FL45_ATTR_COMPARE_LOW ViReal64 s
"Specifies the low comparison value. You cannot query this attribute. \n"
"\n"
"The value you specify for the FL345_ATTR_FUNCTION attribute determines the "
"units of this attribute as shown below.\n"
" \n"
"DC Volts - volts\n"
"AC Volts - volts\n"
"DC Current - amperes\n"
"AC Current - amperes\n"
"2-Wire Resistance - ohms\n"
"AC plus DC Volts - volts\n"
"AC plus DC Current - amperes\n"
"Frequency - hertz\n"
"Diode - volts\n"
"Continuity - volts\n"
"\n"
"Valid range: The value must be within the boundaries (inclusive) of the "
"selected measurement range. The value can be positive or negative.\n"
3 all "Compare High" FL45_ATTR_COMPARE_HIGH ViReal64 s
"Specifies the high comparison value. You cannot query this attribute. \n"
"\n"
"The value you specify for the FL345_ATTR_FUNCTION attribute determines the "
"units of this attribute as shown below.\n"
" \n"
"DC Volts - volts\n"
"AC Volts - volts\n"
"DC Current - amperes\n"
"AC Current - amperes\n"
"2-Wire Resistance - ohms\n"
"AC plus DC Volts - volts\n"
"AC plus DC Current - amperes\n"
"Frequency - hertz\n"
"Diode - volts\n"
"Continuity - volts\n"
"\n"
"Valid range: The value must be within the boundaries (inclusive) of the "
"selected measurement range. The value can be positive or negative.\n"
2 all "Min Max Modifier"
"This group specifies attributes for turning on/off and configuring the "
"min-max modifier.\n"
3 all "Min Max Mode" FL45_ATTR_MIN_MAX_MODE ViInt32 gs attrMinMaxModeRangeTable
"Specifies whether to enable the min-max modifier and whether the min or "
"max value should be displayed. Enabling this modifier automatically "
"disables auto-ranging.\n"
"\n"
"Each time the DMM takes a measurement, it compares the measured value "
"against the minimum and maximum values. If the measured value is less "
"than the minimum value, the DMM saves the measured value as the new "
"minimum value. If the measured value is greater than the maximum value, "
"the DMM saves the measured value as the new maximum value.\n"
"\n"
"Note: \n"
"\n"
"(1) You us the FL45_FetchMinMax function to obtain the minimum and maximum "
"values.\n"
" "
3 all "Min Max Maximum" FL45_ATTR_MIN_MAX_MAX ViReal64 s
"Specifier the value you want to use for the initial maximum value for the "
"min-max modifier. You cannot query this attribute\n"
"\n"
"The value you specifies for the FL45_ATTR_FUNCTION attribute determines "
"the units of this attribute as shown below.\n"
" \n"
"DC Volts - volts\n"
"AC Volts - volts\n"
"DC Current - amperes\n"
"AC Current - amperes\n"
"2-Wire Resistance - ohms\n"
"AC plus DC Volts - volts\n"
"AC plus DC Current - amperes\n"
"Frequency - hertz\n"
"Diode - volts\n"
"Continuity - volts\n"
"\n"
"Valid range: The value must be within the boundaries (inclusive) of the "
"selected measurement range. The value can be positive or negative.\n"
"\n"
"(1) This parameter affects instrument behavior only when you set the "
"Min-Max Mode parameter to FL45_VAL_MIN_MODE or to FL45_VAL_MAX_MODE.\n"
3 all "Min Max Minimum" FL45_ATTR_MIN_MAX_MIN ViReal64 s
"Specifier the value you want to use for the initial minimum value for the "
"min-max modifier. You cannot query this attribute\n"
"\n"
"The value you specifies for the FL45_ATTR_FUNCTION attribute determines "
"the units of this attribute as shown below.\n"
" \n"
"DC Volts - volts\n"
"AC Volts - volts\n"
"DC Current - amperes\n"
"AC Current - amperes\n"
"2-Wire Resistance - ohms\n"
"AC plus DC Volts - volts\n"
"AC plus DC Current - amperes\n"
"Frequency - hertz\n"
"Diode - volts\n"
"Continuity - volts\n"
"\n"
"Valid range: The value must be within the boundaries (inclusive) of the "
"measurement range. The value can be positive or negative.\n"
"\n"
"(1) This parameter affects instrument behavior only when you set the "
"Min-Max Mode parameter to FL45_VAL_MIN_MODE or to FL45_VAL_MAX_MODE.\n"
" "
2 all "Relative Modifier"
"This group specifies the attributes for turning on/off and configuring the "
"relative modifier.\n"
3 all "Relative Enable" FL45_ATTR_RELATIVE_ENABLE ViBoolean gs Boolean_values
"Specifies whether to enable the relative modifier. When the relative "
"modifier is enabled, the DMM returns all measured values relative to a "
"base value. You specify the base value with the FL45_ATTR_RELATIVE_BASE "
"attribute. For example, if the relative base is 15.00 volts, and the "
"present reading is 14.10 volts, the DMM returns -0.90 volts as the result "
"of the FL45_Measure, FL45_Read, and FL45_Fetch functions.\n"
"\n"
"Enabling the modifier automatically disables auto-ranging. When the "
"relative modifier is first enabled, the DMM stores the last valid reading "
"as the relative base. if a last valid reading does not exist, then the "
"relative base is set to 0.0.\n"
3 all "Relative Base" FL45_ATTR_RELATIVE_BASE ViReal64 gs
"Specifies the instruments base value with which the DMM makes relative "
"readings. You cannot access this attribute unless the "
"FL45_ATTR_RELATIVE_MODE is set to VI_TRUE. \n"
"\n"
"The value you specifies for the FL45_ATTR_FUNCTION attribute determines "
"the units of this attribute as shown below.\n"
" \n"
"DC Volts - volts\n"
"AC Volts - volts\n"
"DC Current - amperes\n"
"AC Current - amperes\n"
"2-Wire Resistance - ohms\n"
"AC plus DC Volts - volts\n"
"AC plus DC Current - amperes\n"
"Frequency - hertz\n"
"Diode - volts\n"
"Continuity - volts\n"
"\n"
"Valid range: The value must be within the boundaries (inclusive) of the "
"selected measurement range. The value can be positive or negative.\n"
" "
1 all "Miscellaneous Attributes"
"Additional Attributes."
2 all "ID Query Response" FL45_ATTR_ID_QUERY_RESPONSE ViString g
"Returns the ID Query response string. The instrument driver gets the "
"value of this attribute when you pass VI_TRUE for the ID Query parameter "
"to the FL45_init or FL45_InitWithOptions function.\n"
" "
2 all "OPC Callback Timeout" FL45_ATTR_OPC_TIMEOUT ViInt32 hidden
"This attribute is hidden. The driver uses this attribute internally to "
"set the timeout for the OPC callback.\n"
" "
2 all "FL45_ATTR_MATH_MODIFIERS" FL45_ATTR_MATH_MODIFIERS ViInt32 hidden
""